Patient risk
Let's consider the patient risk implications for different laboratory scenarios that vary in the number of patient specimens examined between QC evaluations and/or the number of QC evaluations performed per day to determine which approach to QC frequency is most critical. For simplicity we will assume that patient testing is spread out evenly across the day and that each lab is applying the same QC rules on the same diagnostic instrument.
Laboratory "A" (Fig. 1A) tests two levels of QC every day and examines an average of 50 patient specimens per day. In the cartoon depictions the horizontal axis is time, the vertical lines represent patient specimens, and the diamonds represent QC evaluations. Laboratory "B" (Fig. 1B) evaluates QC at the same frequency as Lab "A" in terms of time, but Lab "B" examines an average of 300 patient specimens per day. Laboratory "C" (Fig. 1C) tests two levels of QC every eight hours and examines an average of 150 patient specimens daily. Lab "C" evaluates QC at the same frequency as Lab "A" in terms of number of patient specimens between QC evaluations, but at a different frequency in terms of time.
From CLIA's perspective, both labs "A" and "B" are meeting the regulation's minimum requirement by examining two QC samples every 24 hours, while Lab "C" is exceeding the minimum requirements by doing three times as much QC-six QC examinations every day, split into three QC evaluations.
To evaluate the implications of these QC strategy design decisions, let's look at what happens when the test systems malfunction in a non-obvious manner in the middle of the day. There is little risk in an obvious malfunction producing unreliable results; it is the unrecognized malfunction that increases patient risk. A shift in results can be difficult to detect prior to a QC event because the true patient concentrations are unknown. Fig. 2 depicts a malfunction that shifts results higher to the extent that there is a 40% chance that the examination error in an individual patient result is large enough to make the result unreliable (the error exceeds an allowable total error).
In Fig. 2, each red asterisk represents an unreliable patient result that is produced due to the unrecognized malfunction. After the malfunction is detected by a failing QC event, it is corrected and subsequent results are reliable. This example is somewhat optimistic in assuming that the first QC event will detect the malfunction. Depending on the magnitude of the error and the QC rules used, it may take several QC evaluations to detect the malfunction.
When an undetected, persistent malfunction occurs, the number of unreliable patient results produced will be proportional to the number of specimens examined while the test system is malfunctioning. The number of unreliable results produced by Lab "A" is approximately the same as the number of unreliable results produced by Lab "C" (~10 unreliable results), but Lab "B" produced six times as many unreliable results as labs "A" or "C" (~60 unreliable results).
Fig. 3
The number of patient specimens examined between QC evaluations is 50 for labs "A" and "C", but 300 for Lab "B." If Lab "B" examined the same number of patient specimens between QC evaluations as Lab "A" (every 50 patient specimens, or every four hours instead of every 24 hours) there would be a similar outcome in the event of a malfunction (Fig. 3). The number of patient specimens examined between QC evaluations is the critical design factor for managing the risk of producing unreliable patient results in the presence of an undetected malfunction.